AD-4404 Page
121
10.4.
10.4.
10.4.
10.4.
Initializing Parameters
Initializing Parameters
Initializing Parameters
Initializing Parameters
This function initializes the parameters stored in the indicator.
The parameters are stored in the flash memory and backup RAM.
Caution
There are reset functions that require re-calibration of the indicator
Note where the parameters are stored.
Kinds of intialization mode
The menu to initialize parameters is as follows:
[Function] - [Set function] - [SystemMgmt] - [init] - [RAM]
[Code]
[Function]
[CAL]
[All]
Display Kinds
Description
RAM
Initializing RAM
The backup RAM is reset.
Zero point of the gross display, tare value zeroes.
Code
Initializing all codes
All parameters of all codes are reset.
Function
Initializing the
function parameters
Resets parameters of the function list in flash memory.
CAL
Initializing caribration
data
Calibration data is reset in flash memory. If this
function is used, calibrate the indicator. #1
All
Initializing all
parameters
All parameters and calibration data are reset.
Recalibrate and input all parameters. #1
#1 If the switch on the A/D board set to "Disable", it can not be performed.
The Location of the Parameters and Objects of Initialization Mode
Zero
Tare
C
ode w
it
h
out
total
Al
l code
Functi
on
l
is
t
Ca
lib
ra
tio
n
data
Backup RAM
o
o
o
o #2
Location
Flash memory
o #1
o
o
RAM
o
o
Code
o
o
Function
o
CAL
o
Kinds
All
o
o
o
o
o
o
#1 Code data can be stored into flash memory. Refer to " 8.26. Memory Backup"
#2 Total data is always stored in backup RAM.