General Data Format
General Data Format
General Data Format
General Data Format
This format is used for the command mode and jet stream mode.
1 2 3 4 5 6 7 8
9 10 11 12
13
14
15
16
17
18
19
20
21
22 23 24 25 26
27
28
R G R S 0 0 9 9
, 1 2 3
4
5
6
7
,
1
2
3
4
5 6 7 8 9
C
R
L
F
Header
Command is echoed. The echoed command is 4 characters
Code number
The code is 4 characters. A comma "," is appended after this.
Weighing data
Data uses BCD code, is 7 figures and does not include a decimal point.
When data is negative, a minus sign is appended to the head.
A comma "," is appended after this.
Status
The status of weighing data and sequence are indicated at 36 bits.
The status bits synchronizes with control I/O and options.
9 figure is ASCII code. Each lower 4 bits are used.
Each upper 3 bits are fixed to "101".
Terminator
Select a terminator at
Terminator
[
r5 f-7
]
CR or CR + LF
CR:
0D
h, LF:
0A
h
Header
Code number
Weighing data
System Monitor
Terminator
2
0
: Stable
2
1
: Zero
band
2
2
: Full
2
3
: LoLo
1 2
8 9
7
6
5
4
3
2
0
: Lo
2
1
: OK
2
2
: Hi
2
3
: HiHi
2
0
: In operation
2
1
: Conveyor
2
2
: Busy
2
3
:
2
0
: Foreign
material
2
1
: Crush
2
2
: NG
2
3
: Target
count
2
0
:
2
1
:
2
2
: Online
2
3
: In
weighing
sequence
2
0
: Internal reservation
2
1
: Weighing
error
2
2
: Alarm
1
2
3
: Alarm
2
2
0
: Zero
error
2
1
: Out of range
2
2
: Buzzer
2
3
: Tare
entered
2
0
: Center of zero
2
1
: Gross
display
2
2
: Net
display
2
3
: In
hold
display
Internal reservation
Example: 2
3
2
2
2
1
2
0
1
0
1
0
=
A
h
ASCII code =
3A
h.
Upper 4 bits are
3
h always.