AD-4404 Page
127
11.3.
11.3.
11.3.
11.3.
Parameter List
Parameter List
Parameter List
Parameter List
Category address: [Function] - [Set Function] - [General] - [Weighing]
Category
address
symbol
Name Descriptions
Range
and
choices
Default
genf- 1
Display
refresh rate
1: 5 times / second
2: 10 times /second
3: 20 times / second
1 to 3
2
genf- 2
Digital
filter
0: Not used
1: 11 Hz
2: 8.0 Hz
3: 5.6 Hz
4: 4.0 Hz
5: 2.8 Hz
6: 2.0 Hz
7: 1.4 Hz
8: 1.0 Hz
9: 0.7 Hz
Select 3dB band for two
low pass filters.
0 to 99
4 8
genf- 3
Sampling
freqeuncy
divider
Use to decrease the cut-off frequency of
the digital filter.
0 to 10
1
genf- 4
Stability
detection
time
0.0 to
9.9
s
1.0
s
genf- 5
Stability
detection
motion
The detection condition concerning
stablility.
0.0 : Stable at anytime.
0.0 to
9.9
d
2.0
d
genf- 6
Zero range
The range to zero the gross display.
Center of range is zero calibration.
Unit: percentage of weighing capacity.
0 to 30 % 5 %
genf- 7
Zero tracking
time
0.0 to
9.9
s
0.0
s
genf- 8
Zero tracking
motion
The function automatically traces the
weighing deviation at nearly zero point
and keeps zero display of gross display.
If parameter is 0.0, it does not work.
0.0 to
9.9
d
0.0
d
genf- 9
Tare & zero
unstable WGT
When unstable weighing, whether zero
or tare command is used.
0 :Disabled
Relations:
1 :Enabled
genf-4
,
5
0 to 1
1
genf-10
Tare at (-)
gross weight
When negative weghing, whether tare
command is used.
0 :Prohibit tare.
1 : Permission to tare.
0 to 1
1
genf-11
Preset tare
Preset tare
0 :Not used
1 :Use
0 to 1
1
First filter
Second filter
A/D converter