AD-4404 Page
119
10.3.4.
10.3.4.
10.3.4.
10.3.4.
Testing A/D Converter
Testing A/D Converter
Testing A/D Converter
Testing A/D Converter
The A/D converter data is displayed.
When pressing the
ENTER
key, a test voltage can be input to the A/D converter.
10.3.5.
10.3.5.
10.3.5.
10.3.5.
Testing BCD Output of OP
Testing BCD Output of OP
Testing BCD Output of OP
Testing BCD Output of OP----01
01
01
01
One of terminal A and B alternately turns on every one second and in order. The
active terminal displays "1".
10.3.6.
10.3.6.
10.3.6.
10.3.6.
Testing Relay Output of OP
Testing Relay Output of OP
Testing Relay Output of OP
Testing Relay Output of OP----02
02
02
02
A terminal of the relays turns on every one second and in order. The active terminal
displays "1".
10.3.7.
10.3.7.
10.3.7.
10.3.7.
Testing RS
Testing RS
Testing RS
Testing RS----422/485 Interface of OP
422/485 Interface of OP
422/485 Interface of OP
422/485 Interface of OP----03
03
03
03
Each time to press the
ENTER
key, the test data "ST,GS,+0000000kg CR LF" is
output.
The received data is displayed, when data is output.
: CR
: LF
16.001
kg
A/D Count L/C Output(mV/V)
964583 3.103872
Int.Count Test Input
160008 0ff
A/D converter output
Internal count
Loadcell output
Test input
12345
kg
OP-01 000000000000000000
a18 a1
000001000000000000
b18 b1
A1
B1
A18
B18
123
kg
OP-02 0001 00000
10 7 5 1
Output terminal
1 ~ 5
Output terminal
7 ~ 10
Sent data
123
kg
RNET
ST,GS,+0000000kg
0p-03 rcu and 5nd
Received data
Parity:
p
Framing:
f