![AND Check Weighing Indicator AD-4404 Скачать руководство пользователя страница 16](http://html1.mh-extra.com/html/and/check-weighing-indicator-ad-4404/check-weighing-indicator-ad-4404_instruction-manual_2941313016.webp)
Page
14
3.3.
3.3.
3.3.
3.3.
Connecting the Loadcell Cable
Connecting the Loadcell Cable
Connecting the Loadcell Cable
Connecting the Loadcell Cable
Caution
Do not share the loadcell cable with noise-generating devices or power lines,
because the loadcell signal is very sensitive.
We recommend that you use a 6 wire shielded cable to prevent loss of weighing
precision.
If the length of loadcell cable is shorter than 5 m, you may use a 4 wire shielded cable
with terminals 1 & 2 connected together (connected EXC+ and SEN+) and terminals
3 & 4 connected together (connected EXC- and SEN-).
Adaptable Compression Terminal Parts
Use the appropriate compression terminal parts to attach the cables.
Loadcell Output Adjustment for Zero Calibration (Zero Point Adjustment)
If the message "
CERR2
" is displayed, the zero point of zero calibration is too large.
If the message "
CERR3
" is displayed, the zero point of zero calibration is too small.
Use a resistor of more than 50 k
Ω
with low (good) temperature coefficient, when
adding a resistor, to adjust the loadcell output, to the indicator terminals.
In Case of Positive Offset
In Case of Negative Offset
Содержание Check Weighing Indicator AD-4404
Страница 1: ...Check Weighing Indicator WM PD4000489...
Страница 28: ...Page 26 Blank page...