Page
23
5.2.
5.2.
5.2.
5.2.
Digital Span (Calibration without
Digital Span (Calibration without
Digital Span (Calibration without
Digital Span (Calibration without a
a
a
a Mass)
Mass)
Mass)
Mass)
ESC
key
If you want to return to the normal stop mode during the
calibration mode, press the
ESC
key anytime. It is
effective until the last displayed parameter.
Example: zero adjustment only, etc.
ENTER
key When the key is pressed, the procedure stores the
current parameter and proceeds to the next step.
Common Calibration Items
Step 1 Press and hold the
ENTER
key and press the key to display the
menu in the normal stop mode.
Step 2 Press the key twice to select the menu
CAL
.
Press the
ENTER
key to enter the calibration mode.
Step 3 Press the
ENTER
key to enter the menu
CAL
.
Step 4 Select a unit using the numerical keys and press the
ENTER
key
to store it.
Step 5 Select a decimal point using the numerical keys and press the
ENTER
key to store it.
Step 6 Select a minimum division using the numerical keys and press the
ENTER
key to store it.
Step 7 Select a weighing capcity using the numerical keys and press the
ENTER
key to store it.
Step 8 Press the
F1
key to proceed to the next step.
Items for "D
igital
S
pan
"
Step 9 The Zero Point Adjustment
If the zero point value needs adjustment, input it using the
numerical keys and press the
ENTER
key to store it.
If the zero point value does not need adjustment, press the
ENTER
key to proceed the next step.
Step 10 The Span Adjustment
Input the rated capacity of the loadcell unit using the numerical
keys and press the
ENTER
key to store it.
Step 11 Input the sensitivity of the loadcell unit in the unit of mV/V using
the numerical keys and press the
ENTER
key to store it.
Step 12 Press the
ESC
key to return the normal stop mode.
Suggestion
The digital span can be used for trimming of the actual load calibration using a mass.
Содержание Check Weighing Indicator AD-4404