Page
7
Check mode during operation
The monitor mode can confirm system status during operation.
The test mode can test the Input / Output interface.
Even if there is no monitor instrument, the interface can be confirmed.
2.1.
2.1.
2.1.
2.1.
Precaution
Precaution
Precaution
Precautionssss
Before use, confirm the following articles for safe operation.
Grounding the indicator
Ground the indicator. The earth terminal is on the rear panel.
Separate this earth ground line from others, like ground line of a motor, inverter or a
power source. Unless the indicator is grounded, it may result in receiving an electric
shock, cause operation error or catch fire.
Use an adequate power cord
Confirm the AC voltage, current of the power cord and the receptacle type. If the
voltage range of the cord is lower than the power line voltage, it may cause leakage or
catching fire. Use compression terminals to connect the power cord to the rear panel
terminals.
Fuse
The fuse is installed to help prevent the indicator from catching fire.
The indicator is equipped with many safety circuits. Therefore, the fuse is not damaged
in normal operation. If the fuse is damaged, do not replace it, contact your local A&D
dealer. This trouble may have been caused by strong electric discharge.
Splashing
water
The indicator is not water-resistant. When the indicator is mounted to a panel with the
accessory rubber seal, the front panel is equivalent to IP-65.
Flammable
gas
Do not install the indicator where flammable gas is present.
Heat radiation of the indicator
Space out instruments to radiate heat sufficiently.
Use a cooling fan to keep the operating temperature of the indicator within
specifications.
Removing the cover
Disconnect the indicator from the power source before removing the cover to avoid
receiving an electric shock.
Do not touch the internal circuit within 10 seconds after turning off the indicator to
avoid receiving an electric shock.