Page
25
5.4.
5.4.
5.4.
5.4.
Calib
Calib
Calib
Calibration Error
ration Error
ration Error
ration Error
Error Code Error Status and Solution
CERR1
Resolution (Weighing capacity / minimum division) exceeds the limitation.
Increase minimum division or decrease weighing capacity.
CERR2
The initial load (no load output) is larger than 2mV/V.
Check the weighing conveyor, loadcell unit and cable.
CERR3
Negative loadcell output value.
Check the weighing conveyor, loadcell unit and cable.
CERR4
Mass value exceeds the weighing capacity.
Use a mass within the weighing capacity. (Decrease mass value)
CERR5
Mass value is too light for the calibration.
Increase mass value.
CERR6
The loadcell output to be equivalent to the minimum division is too small.
Use a greater minimum division.
CERR7
The polarity of the loadcell output is reversed.
Check the loadcell cable.
CERR8
The mass value of the weighing capacity exceeds 3.2 mV/V.
Confirm the mass and weighing capacity.
CERR9
Gravity acceleration is out of range.
Correct the value within the range of 9.770 ~ 9.835 m/s
2
.
CERR10
Zero output of the loadcell unit is out of range.
Trim the zero output within 0.0 ~ 2.0 mV/V.
CERR11
The loadcell output to be equivalent to minimum division is out of range.
Trim the output within 0.0 ~ 3.2 mV/V.