Chapter 2 System Overview
Part Number: 0160-6411, Rev A, 02/2019
Page 45 of 314
Software Release: cOS 5.00
8. When the system detects that the suction ring has been captured, the Lock panel opens.
As instructed in the Lock panel, adjust the position of the suction ring until all three
indicators (i.e., the vertical green bar on the left; the circular green area over the video;
and the vertical green bar on the right) are within their respective Lock Zones, as shown
in the following figure.
Figure 2.17 Lock Panel
9. When the three indicators are within their respective Lock Zones, press the
(Lock)
I
(on) button on the docking keypad.
10. When the system detects that the lock has been secured, the Verify Fluid panel opens.
As instructed in the Verify Fluid panel, verify that the
LIQUID OPTICS™
Interface is
completely filled with fluid and that no air bubbles are present.
Figure 2.18 Verify Fluid Panel
11. After verifying the fluid, press the FLUID CONFIRMED button to go to the Verify Scan
Screen and initiate
INTEGRAL GUIDANCE
System imaging.
Содержание Catalys OptiMedica
Страница 1: ...Part Number 0160 6411 Rev A 02 2019 Page 1 of 314 Software Release cOS 5 00 Operator Manual ...
Страница 237: ...Part Number 0160 6411 Rev A 02 2019 Page 237 of 314 Software Release cOS 5 00 System Labels and Symbols 282 ...
Страница 238: ...Page 238 of 314 Part Number 0160 6411 Rev A 02 2019 Software Release cOS 5 00 This page was intentionally left blank ...
Страница 287: ...Part Number 0160 6411 Rev A 02 2019 Page 287 of 314 Software Release cOS 5 00 Appendix B Patient Chair Instructions ...
Страница 299: ...Part Number 0160 6411 Rev A 02 2019 Page 299 of 314 Software Release cOS 5 00 Appendix D Software Error Code Table ...