Chapter 4 CATALYS® System Procedure
Part Number: 0160-6411, Rev A, 02/2019
Page 187 of 314
Software Release: cOS 5.00
3. As the suction ring and disposable lens get closer to each other, use your thumb and
pointer finger as a tactile guide to verify that the suction ring and disposable lens are
aligned.
Figure 4.28 Verify by Touch the Alignment of Suction Ring and Disposable Lens
4. When the suction ring and disposable lens are properly mated, release your hand from
the suction ring.
QUICK TIP: CAPTURING THE SUCTION RING
If you do not release the suction ring at this point, your finger may interfere or prevent
capturing the suction ring in the disposable lens.
5. Using the joystick, raise the bed/chair until the suction ring (represented by white/green
arrow on screen) is in the Capture Zone, as shown in the figure on the right.
Figure 4.29 Suction Ring Outside Capture Zone
Figure 4.30 Suction Ring Inside Capture Zone
Use Both Fingers to Verify
Suction Ring and Disposable
Lens Alignment