Chapter 4 CATALYS® System Procedure
Page 198 of 314
Part Number: 0160-6411, Rev A, 02/2019
Software Release: cOS 5.00
7. Review the treatment parameters on the Final Review Screen. If desired, press the BACK
button to return to the Incision Review Screen(s). Otherwise, start laser treatment, as
described in the following section.
Figure 4.42 Final Review Screen and Start Treatment
Performing Laser Treatment
WARNING
Before initiating laser treatment, inspect the images created from the OCT data, surface fits, and
overlaid pattern in both axial and sagittal views, and review the treatment parameters on the
Final Review Screen for accuracy.
QUICK TIP: TREATMENT
Watch the video image during the treatment. Confirm that large bubbles turn into small bubbles
during the capsulotomy, to demonstrate breakthrough of the anterior capsule. You can toggle
the overlay images on and off during the treatment in order to see the bubbles better.
After verifying the treatment parameters on the Final Review Screen, press the laser footswitch
to initiate laser treatment. The Treatment Progress Screen displays.
Содержание Catalys OptiMedica
Страница 1: ...Part Number 0160 6411 Rev A 02 2019 Page 1 of 314 Software Release cOS 5 00 Operator Manual ...
Страница 237: ...Part Number 0160 6411 Rev A 02 2019 Page 237 of 314 Software Release cOS 5 00 System Labels and Symbols 282 ...
Страница 238: ...Page 238 of 314 Part Number 0160 6411 Rev A 02 2019 Software Release cOS 5 00 This page was intentionally left blank ...
Страница 287: ...Part Number 0160 6411 Rev A 02 2019 Page 287 of 314 Software Release cOS 5 00 Appendix B Patient Chair Instructions ...
Страница 299: ...Part Number 0160 6411 Rev A 02 2019 Page 299 of 314 Software Release cOS 5 00 Appendix D Software Error Code Table ...