Appendix A – Chapter 2 Mobile Patient Bed Operation
Part Number: 0160-6411, Rev A, 02/2019
Page 257 of 314
Software Release: cOS 5.00
To activate pivot lock mode, press the
(pivot lock) button on the control pendant. The
button illuminates, and the rolling mechanism on the patient right caster at the front of the bed
locks; the swivel mechanism remains unlocked. The rolling and swivel mechanisms on the
remaining three casters also remain unlocked.
To deactivate pivot lock mode, press the
(pivot lock) button on the control pendant.
Activating/Deactivating Caster Lock Mode
Caster lock mode enables the user to simultaneously lock all four casters to immobilize the bed.
To activate caster lock mode:
• press the
(caster lock) button on the control pendant;
• press the
(caster lock) button on the patient right side of the backrest or the patient
left side of the footrest; or
• press the lock foot pedal on either side of the bed.
When caster lock mode is activated, the caster lock buttons on the control pendant, backrest,
and footrest illuminate.
To deactivate caster lock mode, press any caster lock button on the control pendant, backrest,
or footrest.
Manually Unlocking the Casters
If the Mobile Patient Bed becomes unresponsive and/or the emergency stop button is pressed,
you can manually unlock the casters by rotating the knob on the face of each caster counter-
clockwise.
2.4 Moving the Bed
WARNING
When adjusting or moving the Mobile Patient Bed, always ensure that the patient is properly
positioned within the confines of the bed. Do not allow the patient’s extremities to protrude over
the side of the bed or between the bed rails when adjusting or moving the bed.
Do not move the Mobile Patient Bed over a threshold or other obstacle greater than 10 mm in
height. Doing so may result in user or patient injury.
To minimize vibrations to the patient, pull (do not push) the bed over any threshold.
CAUTION
Do not roll the bed over the AC power cord, control pendant cord, or any
CATALYS®
System cords.
Doing so may permanently damage the cords, requiring replacement.
The Mobile Patient Bed can be moved in the chair or bed position. To move the Mobile Patient
Bed:
Содержание Catalys OptiMedica
Страница 1: ...Part Number 0160 6411 Rev A 02 2019 Page 1 of 314 Software Release cOS 5 00 Operator Manual ...
Страница 237: ...Part Number 0160 6411 Rev A 02 2019 Page 237 of 314 Software Release cOS 5 00 System Labels and Symbols 282 ...
Страница 238: ...Page 238 of 314 Part Number 0160 6411 Rev A 02 2019 Software Release cOS 5 00 This page was intentionally left blank ...
Страница 287: ...Part Number 0160 6411 Rev A 02 2019 Page 287 of 314 Software Release cOS 5 00 Appendix B Patient Chair Instructions ...
Страница 299: ...Part Number 0160 6411 Rev A 02 2019 Page 299 of 314 Software Release cOS 5 00 Appendix D Software Error Code Table ...