Chapter 2 System Overview
Page 38 of 314
Part Number: 0160-6411, Rev A, 02/2019
Software Release: cOS 5.00
Network Ports
The
CATALYS®
System is equipped with a single RJ45 Ethernet connection in the system rear
panel. The Ethernet port conforms to the minimum requirements of:
• Ethernet standard 802.3u - 100BASE-TX, 100BASE-T4, 100BASE-FX
• Fast Ethernet at 100 Mbps (12.5 Mbps) with auto-negotiation
The Ethernet port to which the
CATALYS®
System is connected at the installation site must
conform to these same standards and have active access to the internet via port 80 or 443.
Network Cabling
Ethernet cables connected to the
CATALYS®
System must be Category 6 and meet all the
requirements defined in the IEEE 802.3 standards. Ethernet cables must follow the straight
through design. Cables cannot use the crossover design. Maximum cable length must not exceed
100 m (330 ft). There is no specified minimum length per the IEEE standard.
Setting Up Network Printing and Remote Connectivity (optional)
If desired, you can print treatment reports to a networked printer and/or establish remote
connectivity with
OPTIMEDICA
Service Personnel. The
CATALYS®
System must be configured to
print to the networked printer by
OPTIMEDICA
-trained personnel during system installation. If
the desired printer is changed or a new printer is added to your network, contact
OPTIMEDICA
Service to have the networked printer configured for use with your
CATALYS®
System.
To set up remote connectivity:
1. Connect an Ethernet cable to the RJ45 Network receptacle on the rear of the system
console.
2. Connect the other end of the Ethernet cable to your local network port.
3. When logged in with administrative privileges, press the ENABLE REMOTE SERVICE
button on the Settings Screen.
To disable the network connection, press the DISABLE REMOTE SERVICE button on the Settings
Screen. The network connection is automatically disabled during treatment.
NOTE
Refer to
Printing the Treatment Report(s)
on page
for detailed instructions on printing the
treatment report to a network printer.
Moving the System
The
CATALYS®
System is a delicate medical instrument and is not designed for the user to move
or relocate the system after installation. If the system needs to be moved or relocated within the
installation facility, contact
OPTIMEDICA
. Moving the system must be performed by trained
OPTIMEDICA
personnel.
Trained
OPTIMEDICA
Service Personnel will de-install the system and take the precautions
necessary to ensure a successful move. Once the system has been positioned in its new location,
Содержание Catalys OptiMedica
Страница 1: ...Part Number 0160 6411 Rev A 02 2019 Page 1 of 314 Software Release cOS 5 00 Operator Manual ...
Страница 237: ...Part Number 0160 6411 Rev A 02 2019 Page 237 of 314 Software Release cOS 5 00 System Labels and Symbols 282 ...
Страница 238: ...Page 238 of 314 Part Number 0160 6411 Rev A 02 2019 Software Release cOS 5 00 This page was intentionally left blank ...
Страница 287: ...Part Number 0160 6411 Rev A 02 2019 Page 287 of 314 Software Release cOS 5 00 Appendix B Patient Chair Instructions ...
Страница 299: ...Part Number 0160 6411 Rev A 02 2019 Page 299 of 314 Software Release cOS 5 00 Appendix D Software Error Code Table ...