Appendix D – Software Error Code Table
Page 302 of 314
Part Number: 0160-6411, Rev A, 02/2019
Software Release: cOS 5.00
Clearable Both
04-004
Lateral force on Liquid Optics
Interface exceeds docked limit, patient
automatically undocked
Clearable Both
04-005
Chair sync error, ensure chair is locked in 19° or 0° position before
proceeding
Clearable Both
04-006
Joystick checksum error, contact OptiMedica Service
Clearable Both
04-007
Vacuum on button discrepancy, release button before proceeding
Clearable Both
04-008
Vacuum off button discrepancy, release button before proceeding
Clearable Both
04-009
Capture on button discrepancy, release button before proceeding
Clearable Both
04-010
Capture off button discrepancy, release button before proceeding
Clearable Both
04-011
Lock on button discrepancy, release button before proceeding
Clearable Both
04-012
Lock off button discrepancy, release button before proceeding
Clearable Both
04-013
Vacuum on footswitch discrepancy, release footswitch before
proceeding
Clearable Both
04-014
Vacuum/chair footswitch discrepancy, release footswitch before
proceeding
Clearable Both
04-015
Laser footswitch discrepancy, release footswitch before proceeding
Clearable Both
04-016
Bed disconnected from system, patient automatically undocked
Clearable Both
04-017
Bed emergency stop button pressed, rotate button clockwise until
button is released to reset
Clearable Both
04-018
Bed subsystem error, contact OptiMedica Service
Critical
05-001
ZL #1 wire error, contact OptiMedica Service
Critical
05-002
ZL #2 wire error, contact OptiMedica Service
Critical
05-003
ZL #1 index pulse error, contact OptiMedica Service
Critical
05-004
ZL #2 index pulse error, contact OptiMedica Service
Critical
05-005
ZL mismatch error, contact OptiMedica Service
Critical
05-006
ZL crosswiring error, contact OptiMedica Service
Critical
05-007
Zed #1 wire error, contact OptiMedica Service
Critical
05-008
Zed #2 wire error, contact OptiMedica Service
Critical
05-009
Zed #1 index pulse error, contact OptiMedica Service
Critical
05-010
Zed #2 index pulse error, contact OptiMedica Service
Error
Type
Error
Number
Error
Message
Содержание Catalys OptiMedica
Страница 1: ...Part Number 0160 6411 Rev A 02 2019 Page 1 of 314 Software Release cOS 5 00 Operator Manual ...
Страница 237: ...Part Number 0160 6411 Rev A 02 2019 Page 237 of 314 Software Release cOS 5 00 System Labels and Symbols 282 ...
Страница 238: ...Page 238 of 314 Part Number 0160 6411 Rev A 02 2019 Software Release cOS 5 00 This page was intentionally left blank ...
Страница 287: ...Part Number 0160 6411 Rev A 02 2019 Page 287 of 314 Software Release cOS 5 00 Appendix B Patient Chair Instructions ...
Страница 299: ...Part Number 0160 6411 Rev A 02 2019 Page 299 of 314 Software Release cOS 5 00 Appendix D Software Error Code Table ...