Appendix A – Chapter 3 Mobile Patient Bed Maintenance
Page 272 of 314
Part Number: 0160-6411, Rev A, 02/2019
Software Release: cOS 5.00
Control Pendant Faults
The control pendant icons and buttons flash to indicate various fault conditions.
Table 3.10 Control Pendant Faults, Probable Causes, and Suggestions
Fault
Probable Cause
Suggestion
Pivot lock, caster lock, and
directional lock buttons are
blinking
Caster error
Press the reset button on the base of the bed,
and then calibrate the bed by pressing and
holding the home button until the bed stops
moving (approximately 30 seconds).
Pivot lock, caster lock, and
directional lock buttons are
not functioning as expected
Caster error
Verify that you have pressed the appropriate
button(s) for the desired caster lock mode. If
this does not fix the issue, press the reset but-
ton on the base of the bed, and then calibrate
the bed by pressing and holding the home but-
ton until the bed stops moving (approximately
30 seconds).
Footrest, backrest, seat, and
column buttons are blinking
Emergency stop button
pressed
Rotate the button clockwise to reset.
General error
Press the reset button on the base of the bed,
and then calibrate the bed by pressing and
holding the home button until the bed stops
moving (approximately 30 seconds).
Connect button is blinking
slowly
System is attempting to estab-
lish Bluetooth connection
between Mobile Patient Bed
and
CATALYS®
System
Wait for the Bluetooth connection to be estab-
lished (button no longer flashes but is still illu-
minated).
Connect button is blinking
rapidly
Bluetooth connection
between Mobile Patient Bed
and
CATALYS®
System has
failed
The system automatically attempts to recon-
nect. If automatic reconnection is unsuccessful,
press the connect button to reconnect.
Battery status icon is blinking
green
Battery power is low
Connect the Mobile Patient Bed to an AC power
source after completing the current treatment.
Battery status icon is blinking
red
Battery power is critically low
Connect the Mobile Patient Bed to an AC power
source immediately to prevent power loss.
Battery status icon is solid
red
Zero battery power
Connect the Mobile Patient Bed to an AC power
source to recharge.
Содержание Catalys OptiMedica
Страница 1: ...Part Number 0160 6411 Rev A 02 2019 Page 1 of 314 Software Release cOS 5 00 Operator Manual ...
Страница 237: ...Part Number 0160 6411 Rev A 02 2019 Page 237 of 314 Software Release cOS 5 00 System Labels and Symbols 282 ...
Страница 238: ...Page 238 of 314 Part Number 0160 6411 Rev A 02 2019 Software Release cOS 5 00 This page was intentionally left blank ...
Страница 287: ...Part Number 0160 6411 Rev A 02 2019 Page 287 of 314 Software Release cOS 5 00 Appendix B Patient Chair Instructions ...
Страница 299: ...Part Number 0160 6411 Rev A 02 2019 Page 299 of 314 Software Release cOS 5 00 Appendix D Software Error Code Table ...