Chapter 3 Software Navigation
Page 128 of 314
Part Number: 0160-6411, Rev A, 02/2019
Software Release: cOS 5.00
Figure 3.82 Docking Screen with Lock Panel Open—
Suction Ring/Disposable Lens Assembly Outside Lock Zone
When the green ring appears in the center of the eye image, the suction ring is captured. The
green areas on the three different indicators represent the Lock Zones. The line in the center
indicates the lateral force on the disposable lens. Move the bed/chair in the opposite direction
of the ball at the end of the line (or in the direction of the ball, if the bed/chair direction has
been reversed on the Settings Screen), to reduce the force on the disposable lens, thereby
moving the ball into the Lock Zone.
The arrow at the right indicates the vertical position of the suction ring/disposable lens
assembly. Move the bed/chair up or down to center the suction ring/disposable lens assembly in
the Lock Zone. The arrow at the left indicates the vertical force on the suction ring/disposable
lens assembly, which will also be affected by moving the bed/chair up or down.
Содержание Catalys OptiMedica
Страница 1: ...Part Number 0160 6411 Rev A 02 2019 Page 1 of 314 Software Release cOS 5 00 Operator Manual ...
Страница 237: ...Part Number 0160 6411 Rev A 02 2019 Page 237 of 314 Software Release cOS 5 00 System Labels and Symbols 282 ...
Страница 238: ...Page 238 of 314 Part Number 0160 6411 Rev A 02 2019 Software Release cOS 5 00 This page was intentionally left blank ...
Страница 287: ...Part Number 0160 6411 Rev A 02 2019 Page 287 of 314 Software Release cOS 5 00 Appendix B Patient Chair Instructions ...
Страница 299: ...Part Number 0160 6411 Rev A 02 2019 Page 299 of 314 Software Release cOS 5 00 Appendix D Software Error Code Table ...