Chapter 4 CATALYS® System Procedure
Part Number: 0160-6411, Rev A, 02/2019
Page 185 of 314
Software Release: cOS 5.00
Figure 4.24 Fill the Suction Ring with Saline
6. Rotate the Mobile Patient Bed or patient chair under the laser, taking care not to spill
any fluid and maintaining slight downward pressure on the suction ring with your hand.
NOTE
Properly route the suction ring tubing so it is visible by the surgeon and/or system
operator to observe for fluid movement or possible suction loss, and to ensure that it
does not become entangled, pinched, and/or severed.
7. If using the Mobile Patient Bed, press the
(caster lock) button on the control pendant
to place the bed in caster lock mode.
8. If using the patient chair, latch the chair using the foot-activated chair latching lever
located at the chair base.
Figure 4.25 Location of Chair Latching Lever on Patient Chair
QUICK TIP: LATCHING THE PATIENT CHAIR
Be careful when latching the patient chair in place. Any sudden jerking of the chair can
cause a suction loss. Look at the arrows on the base of the chair and confirm alignment
before gently latching foot lever.
Overflow Ports—It is
Best to Overfill
Chair Latching Lever
Содержание Catalys OptiMedica
Страница 1: ...Part Number 0160 6411 Rev A 02 2019 Page 1 of 314 Software Release cOS 5 00 Operator Manual ...
Страница 237: ...Part Number 0160 6411 Rev A 02 2019 Page 237 of 314 Software Release cOS 5 00 System Labels and Symbols 282 ...
Страница 238: ...Page 238 of 314 Part Number 0160 6411 Rev A 02 2019 Software Release cOS 5 00 This page was intentionally left blank ...
Страница 287: ...Part Number 0160 6411 Rev A 02 2019 Page 287 of 314 Software Release cOS 5 00 Appendix B Patient Chair Instructions ...
Страница 299: ...Part Number 0160 6411 Rev A 02 2019 Page 299 of 314 Software Release cOS 5 00 Appendix D Software Error Code Table ...