Appendix B – Patient Chair Instructions
Part Number: 0160-6411, Rev A, 02/2019
Page 293 of 314
Software Release: cOS 5.00
Patient Chair Specifications
*
OPTIMEDICA
reserves the right to make changes to the device(s) herein. Device(s), therefore, may not
agree in detail with the published design or specifications. All specifications are subject to change with‐
out notice. Contact
OPTIMEDICA
or your local
OPTIMEDICA
representative for information on changes
and new products.
Patient Chair Installation and Setup
The Patient Chair is designed for installation and use inside a sterile operating room or in a non-
sterile pre-op area within an ambulatory surgery center (ASC) or a hospital. Initial installation
and testing will be performed by
OPTIMEDICA
-certified personnel.
During normal use, the Patient Chair operates on electrical power supplied by the
CATALYS®
system.
Environmental Requirements (Operating)
Maximum altitude
7,000 ft (2,134 m)
Temperature range
59 – 89.6 °F (15 – 32 °C)
Maximum humidity
Up to 80% @ 89.6 °F (32 °C) non-condensing
Operating environment
Temperature controlled environment with temperature controlled to
within ±3.6 °F (±2 °C)
Atmospheric pressure
No atmospheric pressure dependency during the operation of the device.
Environmental Requirements (Non-operating)
Maximum altitude
Standard commercial shipping altitude
Temperature range
32 – 131°F (0 – 55°C)
Maximum Humidity
Up to 80% @ 55°C, non-condensing
Physical Characteristics
Length
70.5 in (1.79 m)
Back Width
26.75 in(.68 m)
Seat Width
21.25 in (.54 m)
Chair height
Lowest position: 22 in (56 cm)
Highest position: 26 in (66 cm)
Weight
480 lb (218 kg)
Maximum patient weight
500 lb (227 kg)
Safe Working Load
534 lb (242 kg)
Articulation Range
Backrest inclination
0° to 75°
Swivel range (lockable)
25° right and left from center
Classifications
IEC 60601-1 protection against
electric shock
Type B
IEC 60601-1 patient leakage
Footrest, seat, backrest, headrest, armrests
Содержание Catalys OptiMedica
Страница 1: ...Part Number 0160 6411 Rev A 02 2019 Page 1 of 314 Software Release cOS 5 00 Operator Manual ...
Страница 237: ...Part Number 0160 6411 Rev A 02 2019 Page 237 of 314 Software Release cOS 5 00 System Labels and Symbols 282 ...
Страница 238: ...Page 238 of 314 Part Number 0160 6411 Rev A 02 2019 Software Release cOS 5 00 This page was intentionally left blank ...
Страница 287: ...Part Number 0160 6411 Rev A 02 2019 Page 287 of 314 Software Release cOS 5 00 Appendix B Patient Chair Instructions ...
Страница 299: ...Part Number 0160 6411 Rev A 02 2019 Page 299 of 314 Software Release cOS 5 00 Appendix D Software Error Code Table ...