Software Release: cOS 5.00
Capsulotomy Parameters
Capsulotomy parameters, including cut dimensions, laser settings, and applicable safety
margins, are illustrated in the following figures and summarized in the following tables.
Figure 4.1 Relationships of Key Anatomical Diameters
Concentric, geometric relationship of the pupil (bottom left) and limbus (bottom right) shown in
blue shading; capsulotomy diameter (top left) shown by dotted line within orange shading; and
lens fragmentation diameter (top, right) shown in orange shading.
Figure 4.2 Corneal and Iris Safety Margins
Cross‐sectional view highlighting geometric relationships of lens (shown in red) and respective
safety margins for iris and cornea.
Lens Fragmentation
Diameter 3.0–10.0 mm
Limbal Diameter
Pupil Diameter
Capsulotomy Diameter
2.0–8.0 mm
Central Corneal Thickness
Corneal Safety Margin
500 µm
Cut Depth
200–1000 µm
Iris Safety Margin
500 µm
Anterior Corneal Radius
Posterior Corneal Radius
Содержание Catalys OptiMedica
Страница 1: ...Part Number 0160 6411 Rev A 02 2019 Page 1 of 314 Software Release cOS 5 00 Operator Manual ...
Страница 237: ...Part Number 0160 6411 Rev A 02 2019 Page 237 of 314 Software Release cOS 5 00 System Labels and Symbols 282 ...
Страница 238: ...Page 238 of 314 Part Number 0160 6411 Rev A 02 2019 Software Release cOS 5 00 This page was intentionally left blank ...
Страница 287: ...Part Number 0160 6411 Rev A 02 2019 Page 287 of 314 Software Release cOS 5 00 Appendix B Patient Chair Instructions ...
Страница 299: ...Part Number 0160 6411 Rev A 02 2019 Page 299 of 314 Software Release cOS 5 00 Appendix D Software Error Code Table ...