Appendix A – Chapter 3 Mobile Patient Bed Maintenance
Page 266 of 314
Part Number: 0160-6411, Rev A, 02/2019
Software Release: cOS 5.00
Table 3.3 Guidance and Manufacturer's Declaration: Electromagnetic Immunity
(This table applies to IEC60601-1-2, 4th Edition compliant Mobile Patient Bed manufactured on or after January 1st, 2019)
The Mobile Patient Bed is intended for use in the electromagnetic environment specified below. The customer or
the user of the Mobile Patient Bed should ensure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment:
Guidance
Electrostatic Dis-
charge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood, concrete, or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrical fast tran-
sient/ burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output
lines
±2 kV for power supply
lines
±1 kV for input/output
lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV Differential mode
±2 kV common mode
±1 kV
differential mode
±2 kV
common mode
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, short
interruptions, and
voltage variations
on power supply
input lines
IEC 61000-4-11
Voltage Dips 30% reduc-
tion, 25/30 periods
At 0°
Voltage Dips > 95%
reduction, 0.5 period
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and 315°
Voltage Dips > 95%
reduction, 1 period At 0°
Voltage Interruptions >
95% reduction, 250/300
periods
Voltage Dips 30% reduc-
tion, 25/30 periods
At 0°
Voltage Dips > 95% reduc-
tion, 0.5 period
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
Voltage Dips > 95% reduc-
tion, 1 period At 0°
Voltage Interruptions >
95% reduction, 250/300
periods
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
Mobile Patient Bed requires contin-
ued operation during power mains
interruptions, it is recommended
that the Mobile Patient Bed be
powered from an uninterruptible
power supply or a battery.
Power frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power-frequency magnetic fields
should be at levels characteristic of
a typical location in a typical com-
mercial or hospital environment.
Содержание Catalys OptiMedica
Страница 1: ...Part Number 0160 6411 Rev A 02 2019 Page 1 of 314 Software Release cOS 5 00 Operator Manual ...
Страница 237: ...Part Number 0160 6411 Rev A 02 2019 Page 237 of 314 Software Release cOS 5 00 System Labels and Symbols 282 ...
Страница 238: ...Page 238 of 314 Part Number 0160 6411 Rev A 02 2019 Software Release cOS 5 00 This page was intentionally left blank ...
Страница 287: ...Part Number 0160 6411 Rev A 02 2019 Page 287 of 314 Software Release cOS 5 00 Appendix B Patient Chair Instructions ...
Страница 299: ...Part Number 0160 6411 Rev A 02 2019 Page 299 of 314 Software Release cOS 5 00 Appendix D Software Error Code Table ...