Software Release: cOS 5.00
5.5 System Error Codes, Faults, and Messages
If the system encounters an error, an error code and/or message displays on the screen. There
are four types of errors: critical, clearable, no treat, and warning.
Error Types
Critical Errors
Critical errors cannot be corrected. When a critical error occurs, the system goes to a safe state
and is disabled, and the error screen gives instructions on how to proceed. There are three types
of critical errors: Critical Error – Bed/Chair Active, Critical Error – Bed/Chair Inactive, and Critical
Error – Shutdown.
When a Critical Error – Bed/Chair Active occurs, you have three options:
• Lower the Mobile Patient Bed or patient chair to release the patient
• Press the LOGOUT button to log out of the system and go to the Login Screen
• Press the HOME button to go to the Home Screen
The bed/chair remains active, but it can only be lowered to remove the patient from under the
system; all other bed/chair motions are disabled. If you do not take one of the preceding
actions, the system automatically logs you out after 300 seconds.
NOTE
The Mobile Patient Bed or patient chair can only be lowered while the error screen is displayed. If
you press the HOME or LOGOUT button before lowering the bed/chair, the bed/chair is
completely disabled. To release the patient, you must manually lower the headrest by pulling out
the headrest frame tilt locking lever and pressing down on the headrest frame (if using the
Mobile Patient Bed) or by rotating the headrest vertical adjustment knob (if using the patient
chair).
Figure 5.1 Sample Critical Error – Bed/Chair Active
When a Critical Error – Bed/Chair Inactive occurs, you have two options:
• Press the LOGOUT button to log out of the system and go to the Login Screen
• Press the HOME button to go to the Home Screen
Содержание Catalys OptiMedica
Страница 1: ...Part Number 0160 6411 Rev A 02 2019 Page 1 of 314 Software Release cOS 5 00 Operator Manual ...
Страница 237: ...Part Number 0160 6411 Rev A 02 2019 Page 237 of 314 Software Release cOS 5 00 System Labels and Symbols 282 ...
Страница 238: ...Page 238 of 314 Part Number 0160 6411 Rev A 02 2019 Software Release cOS 5 00 This page was intentionally left blank ...
Страница 287: ...Part Number 0160 6411 Rev A 02 2019 Page 287 of 314 Software Release cOS 5 00 Appendix B Patient Chair Instructions ...
Страница 299: ...Part Number 0160 6411 Rev A 02 2019 Page 299 of 314 Software Release cOS 5 00 Appendix D Software Error Code Table ...