Appendix D – Software Error Code Table
Part Number: 0160-6411, Rev A, 02/2019
Page 301 of 314
Software Release: cOS 5.00
Warning
01-029
Chair unlatched during scan or treat, patient unlocked
Warning
01-030
Chair maintenance required, contact OptiMedica Service
Warning
01-032
Possible fluid/suction loss, check for patient movement
Warning
01-033
Bed connection error, verify correct patient position if docked while
system attempts to reconnect
Clearable Stop
02-001
Vertical force on Liquid Optics
Interface exceeds treatment limit,
reduce force before proceeding
Clearable Stop
02-002
Lateral force on Liquid Optics
Interface exceeds treatment limit,
reduce force before proceeding
Clearable Stop
02-003
Door interlock disconnected or open, close door or replace door
interlock before proceeding
Clearable Stop
02-004
Abort button pressed
Clearable Stop
02-005
Laser footswitch lifted during treatment
Clearable Stop
02-006
Possible fluid/suction loss, check for patient movement
Clearable Stop
02-007
Video frame rate too low for treatment, verify Liquid Optics
Interface
is correctly docked and patient is not moving before proceeding
Clearable Machine
03-001
Capture line pressure low, inspect Liquid Optics
Interface and fluid
catchment before attempting to dock
Clearable Machine
03-002
Capture lost, inspect Liquid Optics
Interface and fluid catchment
before attempting to re-dock
Clearable Machine
03-003
Disposable lens detection failure, ensure disposable lens is correctly
installed before proceeding
Clearable Machine
03-004
Disposable lens out of position, ensure disposable lens is correctly
installed before proceeding
Clearable Machine
03-005
Lock not fully engaged, ensure vertical position indicator is centered
in Lock Zone when re-docking
Clearable Both
04-001
Vacuum line pressure low, inspect Liquid Optics
Interface and fluid
catchment before attempting to dock
Clearable Both
04-002
Vacuum lost, inspect Liquid Optics
Interface and fluid catchment
before attempting to re-dock
Clearable Both
04-003
Vertical force on Liquid Optics
Interface exceeds docked limit, patient
automatically undocked
Error
Type
Error
Number
Error
Message
Содержание Catalys OptiMedica
Страница 1: ...Part Number 0160 6411 Rev A 02 2019 Page 1 of 314 Software Release cOS 5 00 Operator Manual ...
Страница 237: ...Part Number 0160 6411 Rev A 02 2019 Page 237 of 314 Software Release cOS 5 00 System Labels and Symbols 282 ...
Страница 238: ...Page 238 of 314 Part Number 0160 6411 Rev A 02 2019 Software Release cOS 5 00 This page was intentionally left blank ...
Страница 287: ...Part Number 0160 6411 Rev A 02 2019 Page 287 of 314 Software Release cOS 5 00 Appendix B Patient Chair Instructions ...
Страница 299: ...Part Number 0160 6411 Rev A 02 2019 Page 299 of 314 Software Release cOS 5 00 Appendix D Software Error Code Table ...