5-6 | Thermox WDG-1200/1210 Insitu Probes
(d) thermocouple fault and heater above maximum limit 40.00.
The analyzer is measuring an unexpected voltage from the probe thermocouple.
cause
element
action
1. thermocouple open circuit.
Probe
Check probe wiring.
2. thermocouple wired in reverse polarity.
Wiring
Check connections.
3. thermocouple failed (very unlikely event). Probe
Replace heater -
thermocouple assembly.
(e) cold junction temperature fault 600.0.
The analyzer is reading an unexpected voltage from starting, or correctly complet-
ing, an auto calibration cycle.
cause
element
action
1. sensor not connected or
incorrectly wired.
Probe
Check probe wiring.
2. Probe cold junction
temperature too high.
Probe
Re-site probe or cool connection box.
3. sensor failed.
Probe
Replace sensor diode.
(f) Failed in calibration 7000.
A fault on the system has prevented the analyzer from starting, or correctly complet-
ing, an automatic calibration.
cause
element
action
1. Zero gas pressure not
detected.
ACu
Ensure zero cal gas is connected and
turned on.
2. Any previous fault
see appropriate fault section.
Содержание Thermox WDG-1200 Insitu
Страница 9: ...ix...
Страница 15: ...1 6 Thermox WDG 1200 1210 Insitu Probes This page intentionally left blank...
Страница 19: ...2 4 Thermox WDG 1200 1210 Insitu Probes This page intentionally left blank...
Страница 86: ...Maintenance and Troubleshooting 5 13 This page intentionally left blank...
Страница 95: ...Service and Parts 6 9 Figure 6 4 Cell clip with attached cell leads Figure 6 5 Cell and retainer clip placement...
Страница 98: ...6 12 Thermox WDG 1200 1210 Insitu Probes This page intentionally left blank...
Страница 100: ...A 2 Thermox WDG 1200 1210 Insitu Probes...