4-14 | Thermox WDG-1200/1210 Insitu Probes
Fault list (diAg)
To display the fault list, press
enter
when the
diagnos-
tics
screen is on view. Each fault is represented by a
particular number and can be identified using the
Fault
Table in Chapter 5
. Multiple faults can be displayed by
adding the fault numbers together. The fault status is
transmitted along the Modbus.
OPTIONS
View faults
RANGE
NOT APPLICABLE
DEFAULT
NOT APPLICABLE
Cell EMF (diAg
↵
CEL)
Press
enter
to display the raw signal from the cell
which is measured in mV.
OPTIONS
Display cell EMF
RANGE
NOT APPLICABLE
DEFAULT
NOT APPLICABLE
Cell temperature (diAg
↵
CEL
↵
CELt)
Press
enter
to display the temperature of the cell. It is
the actual temperature of the cell measured indepen-
dently of the heater.
OPTIONS
Display cell temperature
RANGE
NOT APPLICABLE
DEFAULT
NOT APPLICABLE
Cold junction temperature (diAg
↵
CEL
↵
CJC)
Press
enter
to display the temperature at the cold
junction of the thermocouple.
OPTIONS
Display cold junction temperature
RANGE
NOT APPLICABLE
DEFAULT
NOT APPLICABLE
Cell resistance (diAg
↵
rES)
The cell resistance gives some indication of the cell
condition. Values up to 1000 Ohms are normal. Often
high values are caused by poor continuity of the contacts
at the cell. This can be remedied by cleaning. If this does
not improve, then it may indicate that a cell change will be
needed soon. The cell resistance measurement cannot
function at oxygen concentrations above 15% O
2
.
OPTIONS
Display cell resistance
RANGE
NOT APPLICABLE
DEFAULT
NOT APPLICABLE
Содержание Thermox WDG-1200 Insitu
Страница 9: ...ix...
Страница 15: ...1 6 Thermox WDG 1200 1210 Insitu Probes This page intentionally left blank...
Страница 19: ...2 4 Thermox WDG 1200 1210 Insitu Probes This page intentionally left blank...
Страница 86: ...Maintenance and Troubleshooting 5 13 This page intentionally left blank...
Страница 95: ...Service and Parts 6 9 Figure 6 4 Cell clip with attached cell leads Figure 6 5 Cell and retainer clip placement...
Страница 98: ...6 12 Thermox WDG 1200 1210 Insitu Probes This page intentionally left blank...
Страница 100: ...A 2 Thermox WDG 1200 1210 Insitu Probes...