User Interface / Calibration
|
4-19
analog output settings
All the options necessary to set the analyzer outputs can be accessed using the
Analog Outputs
menu. The
Analog Outputs
menu consists of five sub-menus:
• Zero
• Span
• Damping
• Base
• Track.
To access these sub-menus, from the
Analog Outputs
display, press
Enter
. To
scroll through the menu, press
F1
. Each option within the sub-menu is set by
pressing
Enter
, selecting the appropriate value using the
UP
and
DOWN
keys,
and pressing
Enter
again. The value will be set and the display scrolls on to the
next option.
Zero reading
This is the reading to be output when the analyzer measures at the zero cali-
brated value. To set the reading:
Press
enter
Select a value using the
UP
and
doWn
keys
Press
enter
again.
span reading
This is the reading to be output when the analyzer measures at the span cali-
brated value. To set the reading:
Press
enter
Select a value using the
UP
and
doWn
keys
Press
enter
again.
damping
This is a smoothing facility for the output to stabilize the reading under noisy
signal conditions. To apply the setting:
Press
enter
Select a value using the
UP
and
doWn
keys
Press
enter
again.
base
The output range required determines the base selected. Make a selection
ranging from 0-10 mA to 4-20 mA by pressing
enter
, selecting the preferred
option using the
UP
and
doWn
keys, and pressing
enter
again.
Содержание Thermox WDG-1200 Insitu
Страница 9: ...ix...
Страница 15: ...1 6 Thermox WDG 1200 1210 Insitu Probes This page intentionally left blank...
Страница 19: ...2 4 Thermox WDG 1200 1210 Insitu Probes This page intentionally left blank...
Страница 86: ...Maintenance and Troubleshooting 5 13 This page intentionally left blank...
Страница 95: ...Service and Parts 6 9 Figure 6 4 Cell clip with attached cell leads Figure 6 5 Cell and retainer clip placement...
Страница 98: ...6 12 Thermox WDG 1200 1210 Insitu Probes This page intentionally left blank...
Страница 100: ...A 2 Thermox WDG 1200 1210 Insitu Probes...