Overview
|
1-3
output signal.
Because of the high operating temperature of the cell, combustible
gases that are present may burn. When this occurs, the cell will
generate high millivolts and cause the display to indicate less oxygen
than is actually in the gas (net oxygen content).
Figure 1-1.
Zirconium oxide cell principle of operation.
Three Simple Systems in the Analyzer
• The Mechanical System: The inner and outer pipes, mounting flange, and
terminal case.
• The Measuring System: The sensing cell, interconnecting wiring, and inte-
grated user interface.
• The Temperature System: The electrical zirconium oxide cell heater with
internal type “K” thermocouple, and the control unit. The sensing cell of the
analyzer is operated at a constant temperature. Heat is supplied to the cell by
an electrically powered internal heater. Inside the heater is a thermocouple used
in the cell temperature control circuit.