User Interface / Calibration
|
4-17
output setup
system outputs
The current loop output can provide one of two output functions:
Track
Measurement output is tracked during normal reading and calibration
operations.
Hold
Measurement output is held at the last reading prior to calibration, and
resumes after the final calibration settling period.
Output Held
Setting Time
ACTIVATED
ACTIVATED
ACTIVATED
DEACTIVATED
DEACTIVATED
DEACTIVATED
Analog Outputs Set to ‘HOLD’
(’NO’ selected)
Analog Outputs Set to ‘TRACK’
(’YES’ selected)
Span Level
Zero Level
Calibration in Progress Relay
During Automatic Calibration
Calibration in Progress Relay
During Span Calibration
Calibration in Progress Relay
During Zero Calibration