3-2 | Thermox WDG-1200/1210 Insitu Probes
Mechanical Installation
• Location
• Installing the Filter/Flame Arrestor (optional)
• Mounting the Sensor
• Calibration Setup
location
Observe the following guidelines when selecting an analyzer instal-
lation location.
The installation location should be free from excessive vibration. The ambi-
ent temperature must be within the limits listed in the specifications chapter in
this manual. If the ambient temperature is outside the specified limits or the
vibration is excessive, please contact Thermox Sales or Service Department at
412.828.9040. There are special options to address ambient temperatures out-
side the listed specifications and special-mounting solutions can be provided for
excessive vibration installation locations.
Handle the probe with care,
do not drop the probe
. The probe contains fragile
ceramic components which are easily damaged. The probe should be removed
from the stack/flue when high pressure cleaning with a water hose, as this can
crack the zirconia cell.
Select a suitable accessible position where the probe is located in the main stream
of the flue gas. Flue gas temperatures must be in the range of 20°C/68°F to
800°C/1472°F.
Avoid positions where obstructions or bends may impede the flow of gas or
prevent insertion or subsequent removal of the probe.
Sufficient clearance must be allowed for installation and removal of the probe. A
minimum clearance of the probe length plus an extra 25mm/ 1” from the boiler
flange or stand-off is suggested.
If excessive dust flow is likely, fit a deflector plate with its apex facing the flow of
flue gas.
Содержание Thermox WDG-1200 Insitu
Страница 9: ...ix...
Страница 15: ...1 6 Thermox WDG 1200 1210 Insitu Probes This page intentionally left blank...
Страница 19: ...2 4 Thermox WDG 1200 1210 Insitu Probes This page intentionally left blank...
Страница 86: ...Maintenance and Troubleshooting 5 13 This page intentionally left blank...
Страница 95: ...Service and Parts 6 9 Figure 6 4 Cell clip with attached cell leads Figure 6 5 Cell and retainer clip placement...
Страница 98: ...6 12 Thermox WDG 1200 1210 Insitu Probes This page intentionally left blank...
Страница 100: ...A 2 Thermox WDG 1200 1210 Insitu Probes...