4-8 | Thermox WDG-1200/1210 Insitu Probes
Span gas concentration (CAL
↵
COnF
↵
SgAS)
To set the concentration of the gas used to calibrate
the span of the analyzer:
Press
enter
Select a value using the
UP
and
doWn
keys
Press
enter
again.
Use a concentration of 20.9% O
2
.
OPTIONS
Set concentration of span
calibration gas
RANGE
5 to 25 % O
2
DEFAULT
20.90 % O
2
Zero gas concentration (CAL
↵
COnF
↵
ZgAS)
To set the concentration of the zero calibration gas:
Press
enter
Select a value using the
UP
and
doWn
keys
Press
enter
again.
do not use a concentration of 0% o
2
in n
2
.
It is not possible to calibrate on 0% oxygen, therefore
the recommended minimum concentration is 0.1% O
2
in N
2
.
OPTIONS
Set concentration of zero
calibration gas
RANGE
0.1 to 18 % O
2
in nitrogen
DEFAULT
1 % O
2
in nitrogen
Calibration time interval (CAL
↵
COnF
↵
Cti)
To set the time between calibrations:
Press
enter
Select a value using the
UP
and
doWn
keys
Press
enter
again.
e.g., 24hrs = daily calibration
OPTIONS
Set calibration time interval
RANGE
9999 to 0 hours
DEFAULT
0 hours
Time to next calibration (CAL
↵
COnF
↵
ttnC)
To set the time the analyzer will wait until the first cali-
bration is performed:
Press
enter
Select a value using the
UP
and
doWn
keys
Press
enter
again.
e.g., 120 mins = two hours time
OPTIONS
Set time to next calibration
RANGE
0 to 9999 mins
DEFAULT
0 mins (OFF)
Содержание Thermox WDG-1200 Insitu
Страница 9: ...ix...
Страница 15: ...1 6 Thermox WDG 1200 1210 Insitu Probes This page intentionally left blank...
Страница 19: ...2 4 Thermox WDG 1200 1210 Insitu Probes This page intentionally left blank...
Страница 86: ...Maintenance and Troubleshooting 5 13 This page intentionally left blank...
Страница 95: ...Service and Parts 6 9 Figure 6 4 Cell clip with attached cell leads Figure 6 5 Cell and retainer clip placement...
Страница 98: ...6 12 Thermox WDG 1200 1210 Insitu Probes This page intentionally left blank...
Страница 100: ...A 2 Thermox WDG 1200 1210 Insitu Probes...