3-14 | Thermox WDG-1200/1210 Insitu Probes
Cable recommendations - WdG-1200 Insitu
WdG-1200 Insitu Cable recommendations
ref.
description
Max. o.d.
mm
no. of
Cores
Core
size mm
2
Core size
strands/dia
Core
size awg
shielded
1
Mains Supply
11
3
0.5
16/0.2
22
No
2
Calibration
13
6
0.2
7/0.2
24
Yes
3
Outputs/Relays
15
17
0.2
7/0.2
24
Yes
1
2
3
Cables 1 and 3 are supplied with the analyzer. Cable 2 is supplied with the re-
mote calibration unit.
Cable recommendations - WdG-1210 Insitu
WdG-1210 Insitu Cable recommendations
ref.
description
Max. o.d.
mm
no. of
Cores
Core
size mm
2
Core size
strands/dia
Core
size awg
shielded
1
Mains Supply
11
3
0.5
24/0.2
18
No
2
Calibration
13
6
0.2
7/0.2
24
Yes
3
Outputs/Relays
15
17
0.2
7/0.2
24
Yes
4
Probe Power
11
3
0.8
24/0.2
18
Yes
5
Interconnect
13
6
0.5
16/0.2
22
Yes
Cables 1, 3, 4 and 5 are supplied with the analyzer. Cable 2 is supplied with the
remote calibration unit.
Содержание Thermox WDG-1200 Insitu
Страница 9: ...ix...
Страница 15: ...1 6 Thermox WDG 1200 1210 Insitu Probes This page intentionally left blank...
Страница 19: ...2 4 Thermox WDG 1200 1210 Insitu Probes This page intentionally left blank...
Страница 86: ...Maintenance and Troubleshooting 5 13 This page intentionally left blank...
Страница 95: ...Service and Parts 6 9 Figure 6 4 Cell clip with attached cell leads Figure 6 5 Cell and retainer clip placement...
Страница 98: ...6 12 Thermox WDG 1200 1210 Insitu Probes This page intentionally left blank...
Страница 100: ...A 2 Thermox WDG 1200 1210 Insitu Probes...