User Interface / Calibration
|
4-29
automatic calibration
The
Automatic Calibration
sub-menu has only one option. Access this option
by pressing
Enter
from the
Auto Calibration
display. The
Auto Calibration
option is selected by pressing
Enter
, selecting
YES
using the
UP
key and press-
ing
Enter
again.
▼
automatic calibration sub-Menu
If selected, the analyzer will perform a zero calibration first and then a span
calibration. The calibration gases will be allowed to flow automatically from the
calibration gas unit. To bypass this option, press
f1
.
Check that the zero and span gases are turned on and regulated at 10
PSI (0.7 bar) nominal at the gas bottles.
remote triggering
Analyzer calibration can be triggered remotely from the plant control system.
When triggered, the automatic calibration sequence will begin and calibration
gas will be pumped automatically from the remote calibration unit.
NOTE: Check to ensure the serial port is wired correctly.