Service and Parts
|
6-3
6. Slide the inner probe assembly out of the outer protective tube by pulling on
its handle.
Process under Pressure
If the process is under pressure, place a blind plate and gasket over the flange
hole (blind plate and gasket included in insitu accessory kit).
• Place the blind plate and gasket over the 8-32 stud on top of the flange hole.
• Insert the two 8-32 x 1/4” long screws from the insitu accessory kit into the
two remaining holes. Tighten using a flat blade screwdriver.
• Place the 8-32 kep nut removed in Step 4 over the 8-32 stud.
• Using a 11/32” nut driver. Tighten the kep nut.
If the inner probe assembly does not slide out freely, remove the 10-32 screw
from the sheet metal plate and insert a longer screw (thumb screw provided in in-
situ accessory kit) into this threaded hole to force the inner probe assembly loose
(be sure to remove this longer screw once the probe assembly is loose).
After removing the inner probe assembly, inspect the gasket for dam-
age. If damaged, replace the gasket.
NOTE