4-12 | Thermox WDG-1200/1210 Insitu Probes
Zero setting (AO
↵
ZEro)
To set the minimum O
2
concentration of the
analog output
range
:
Press
enter
Select a value using the
UP
and
doWn
keys
Press
enter
again.
OPTIONS
Set zero reading output
RANGE
0 to 20 %
DEFAULT
0 %
Analog outputs span setting (AO
↵
SPAn)
To set the maximum O
2
concentration of the
analog output
range
:
Press
enter
Select a value using the
UP
and
doWn
keys
Press
enter
again.
OPTIONS
Set span reading output
RANGE
1 % to 25 %
DEFAULT
20.90 %
Analog outputs damping setting (AO
↵
dPng)
To apply a smoothing facility for the output to stabilize the
reading under noisy signal conditions:
Press
enter
Select a value using the
UP
and
doWn
keys
Press
enter
again.
OPTIONS
Setting the damping rate
RANGE
0 to 999.0 sec
DEFAULT
10.00 sec
Base setting (AO
↵
bASE)
The output range required will determine the base selected.
Ranging from 0-10mA to 4-20mA. To make a selection:
Press
enter
Select the preferred option using the
UP
and
doWn
keys
Press
enter
again.
OPTIONS
[0-10][2-10][4-10]
[0-20][2-20][4-20]
RANGE
NOT APPLICABLE
DEFAULT
4-20
Outputs track setting (AO
↵
t
rAc)
The analog output can be set to “track” the O
2
levels or
“hold” the last reading during calibration. To set track or hold:
Press
enter
Select
Yes
(track) or
no
(hold) using the
UP
and
doWn
keys
Press
enter
again.
.
OPTIONS
[NO][YES]
RANGE
NOT APPLICABLE
DEFAULT
NO
Содержание Thermox WDG-1200 Insitu
Страница 9: ...ix...
Страница 15: ...1 6 Thermox WDG 1200 1210 Insitu Probes This page intentionally left blank...
Страница 19: ...2 4 Thermox WDG 1200 1210 Insitu Probes This page intentionally left blank...
Страница 86: ...Maintenance and Troubleshooting 5 13 This page intentionally left blank...
Страница 95: ...Service and Parts 6 9 Figure 6 4 Cell clip with attached cell leads Figure 6 5 Cell and retainer clip placement...
Страница 98: ...6 12 Thermox WDG 1200 1210 Insitu Probes This page intentionally left blank...
Страница 100: ...A 2 Thermox WDG 1200 1210 Insitu Probes...