User Interface / Calibration
|
4-9
Calibration settling time (CAL
↵
COnF
↵
SEtL)
The period required for the calibration gas to stabilize is
normally two minutes. If using long lengths of piping the
time may need to be extended. To enter the settling time:
Press
enter
Select a value using the
UP
and
doWn
keys
Press
enter
again.
OPTIONS
Set time for calibration read-
ings to settle
RANGE
30 to 1600 sec
DEFAULT
120 sec
Automatic calibration (CAL
↵
AutO)
To perform an automatic calibration press
enter
. The
analyzer will perform the calibration and then return to
the O
2
display.
OPTIONS
Perform automatic calibration
RANGE
NOT APPLICABLE
DEFAULT
NOT APPLICABLE
Zero calibrati/on (CAL
↵
SCAL
↵
ZEro)
A zero calibration is performed manually by pressing
enter
. If the reading shown is inaccurate, press
enter
again. If the reading is accurate, the zero calibration
can be bypassed by pressing
f1
.
OPTIONS
Perform zero calibration
RANGE
NOT APPLICABLE
DEFAULT
NOT APPLICABLE
Manual span calibration (CAL
↵
SCAL
↵
SPAn)
A span calibration is performed manually by pressing
enter
. If the reading shown is inaccurate press
enter
again. If the reading is accurate, the span calibration
can be bypassed by pressing
f1
.
OPTIONS
Perform span calibration
RANGE
NOT APPLICABLE
DEFAULT
NOT APPLICABLE
Содержание Thermox WDG-1200 Insitu
Страница 9: ...ix...
Страница 15: ...1 6 Thermox WDG 1200 1210 Insitu Probes This page intentionally left blank...
Страница 19: ...2 4 Thermox WDG 1200 1210 Insitu Probes This page intentionally left blank...
Страница 86: ...Maintenance and Troubleshooting 5 13 This page intentionally left blank...
Страница 95: ...Service and Parts 6 9 Figure 6 4 Cell clip with attached cell leads Figure 6 5 Cell and retainer clip placement...
Страница 98: ...6 12 Thermox WDG 1200 1210 Insitu Probes This page intentionally left blank...
Страница 100: ...A 2 Thermox WDG 1200 1210 Insitu Probes...