User Interface / Calibration
|
4-31
▼
span calibration
The span calibration reading is checked by connecting the calibration gas via
the calibration gas inlet on the probe and pressing
enter
. The reading displayed
will need approximately 120 seconds to settle.
NOTE
Recommended span calibration gas concentration is 20.90% O
2
(i.e., ambient air).
▼
span calibration reading
Calibration is necessary if the reading displayed is incorrect (i.e. not equal to the
span calibration gas concentration). It is performed by holding
enter
for two sec-
onds. If the reading displayed is correct, the option can be bypassed by press-
ing
f1
.
Содержание Thermox WDG-1200 Insitu
Страница 9: ...ix...
Страница 15: ...1 6 Thermox WDG 1200 1210 Insitu Probes This page intentionally left blank...
Страница 19: ...2 4 Thermox WDG 1200 1210 Insitu Probes This page intentionally left blank...
Страница 86: ...Maintenance and Troubleshooting 5 13 This page intentionally left blank...
Страница 95: ...Service and Parts 6 9 Figure 6 4 Cell clip with attached cell leads Figure 6 5 Cell and retainer clip placement...
Страница 98: ...6 12 Thermox WDG 1200 1210 Insitu Probes This page intentionally left blank...
Страница 100: ...A 2 Thermox WDG 1200 1210 Insitu Probes...