Maintenance and Troubleshooting
|
5-3
Fault identification numbers
Each fault is identified by a specific number as follows:
(a) Heater temperature out of the control band (±5°C)
00.02
(b) span coefficient out of range
00.10
(c) Zero coefficient out of range
3.00
(d) thermocouple fault and heater above maximum limit
40.00
(e) Cold junction temperature fault
600.0
(f)
Failed in calibration
7000
The Fault Identification Number may represent a multiple fault, in which case the
individual identification numbers are added together. Use the table below as a quick
reference for faults listed above.
identification
Number
Fault
identification
Number
Fault
00.02
a
600.0
e
00.10
b
600.1
b+e
00.12
a+b
603.0
c+e
03.00
c
603.1
b+c+e
03.02
a+c
640.0
d+e
03.10
a+b
640.1
b+d+e
03.12
a+b+c
643.0
c+d+e
40.00
d
643.1
b+c+d+e
40.02
a+d
7000
f
40.10
b+d
7003
c+f
40.12
a+b+c
7040
d+f
43.00
c+d
7043
c+d+f
43.02
a+d+c
7600
e+f
43.10
b+c+d
7603
c+e+f
43.12
a+b+c+d
7643
c+d+e+f
Table 1. Fault Identification Table
A guide to the cause of faults and actions to take, is listed in the next section.
Содержание Thermox WDG-1200 Insitu
Страница 9: ...ix...
Страница 15: ...1 6 Thermox WDG 1200 1210 Insitu Probes This page intentionally left blank...
Страница 19: ...2 4 Thermox WDG 1200 1210 Insitu Probes This page intentionally left blank...
Страница 86: ...Maintenance and Troubleshooting 5 13 This page intentionally left blank...
Страница 95: ...Service and Parts 6 9 Figure 6 4 Cell clip with attached cell leads Figure 6 5 Cell and retainer clip placement...
Страница 98: ...6 12 Thermox WDG 1200 1210 Insitu Probes This page intentionally left blank...
Страница 100: ...A 2 Thermox WDG 1200 1210 Insitu Probes...