Maintenance and Troubleshooting
|
5-1
maintenance and troubleshooting
The operations in this chapter should be performed only by qualified
service personnel with knowledge of electrical safety techniques. There
are no operator-serviceable components inside the WDG-1200/1210
Insitu probe. Never service the probe unless power has been removed
from the probe and it has been allowed to cool for one hour.
identifying Faults
Faults are always indicated by the illumination of the
System Fault
LED and the
activation of the fault relay. Fault status will also be transmitted along the Modbus.
Most faults can be found on the
Fault
menu; specific problems may be referred to
any of the Thermox sales/service offices for technical advice or assistance.
Always quote the analyzer type and serial number when requesting
technical assistance.