1-4 | Thermox WDG-1200/1210 Insitu Probes
Problems to Avoid
These are some errors to avoid. If they are avoided, your analyzer will operate
with a minimum of maintenance and troubleshooting.
• Do not use pipe dope or any other contaminant that gives off combustible
vapor which may cause erroneous measurements on any joints of the sample
tubing.
• Calibration gases must contain oxygen. Do not use pure nitrogen or any other
anaerobic gas, or a zero gas.
• Do not use calibration gases if they contain a mixture of oxygen and combus-
tibles.
• Do not handle the cell or heater (white end) excessively. Do not try to clean
the cell. When handling, grasp by touching the seal fitting.
Cautions to be observed:
• Turn off the power to the probe when working on the probe.
• All the parts inside the probe are hot and will cause severe burns if touched.
Use gloves, wrenches, and care when working inside the sensor.