4-30 | Thermox WDG-1200/1210 Insitu Probes
Manual calibration
The
Manual Calibration
sub-menu has two options:
• Zero
• Span Calibration
These options are accessed by pressing
Enter
from the
Manual Calibration
display. Both options are initiated by connecting the appropriate calibration gas,
pressing
Enter
, waiting for the reading to settle, and then pressing
Enter
to con-
tinue, or pressing
F1
to bypass the option.
To avoid possible errors which can be introduced by separate zero
and span calibration, the zero and span MUST be calibrated to-
gether.
▼
Manual calibration
Zero and span calibration are performed manually from this sub-menu
.
Set all configuration options before performing calibrations.
▼
Zero calibration
Supply calibration gases to the probe using the calibration gas inlet at a steady
flow of 1.4 l/min (3 ft
3
/hour) to 2.8 l/min. (6 ft
3
/hour), and then press
enter
. The
reading displayed will need approximately 120 seconds to settle.
NOTE
Recommended zero calibration gas concentration is 1% O
2
in nitrogen.
▼
Zero calibration reading
Calibration is necessary if the reading displayed is incorrect (i.e., not equal to the
zero calibration gas concentration). Calibration is performed by holding down
enter
for two seconds. The option can be bypassed by pressing
f1
.
NOTE
NOTE
Содержание Thermox WDG-1200 Insitu
Страница 9: ...ix...
Страница 15: ...1 6 Thermox WDG 1200 1210 Insitu Probes This page intentionally left blank...
Страница 19: ...2 4 Thermox WDG 1200 1210 Insitu Probes This page intentionally left blank...
Страница 86: ...Maintenance and Troubleshooting 5 13 This page intentionally left blank...
Страница 95: ...Service and Parts 6 9 Figure 6 4 Cell clip with attached cell leads Figure 6 5 Cell and retainer clip placement...
Страница 98: ...6 12 Thermox WDG 1200 1210 Insitu Probes This page intentionally left blank...
Страница 100: ...A 2 Thermox WDG 1200 1210 Insitu Probes...