Installation and Start-up
|
3-5
Mounting the sensor
Ambient temperature in the controller area should be less than
160° F (70°C).
Mount the probe and its flange to the flange on the process (see Figure
3-2a and 3-2b). Insulate the pipe nipple with at least 1” (2.54 cm) thick weather-
proof insulation. If the pipe nipple is over 4” (10.16 cm) long, it should be heat
traced.
Figure 3-2a. WdG-1200 Insitu sensor mounting.
6.02
153
5.98
152
1.08
27.43
1.08
27.43
1/2"NPSF
2 PLACES
CL
4.39
111.55
2.59
65.91
3.50
89
.24
6
1.04
26.42
CL
2"NPT (M)
Figure 3-2b. WdG-1210 Insitu sensor mounting.