Calibration | 5-3
Overview
The Calibrate key allows you to do the following (see Figure 5-2) :
• Initiate an automatic, remote, or manual calibration or verification.
Although automatic calibrations can be set to run at predetermined
frequencies, you can also force the system into an automatic calibra-
tion (Initiate Cal and Initiate Verify).
• View previously collected calibration or verify data (Cal/Verify Data).
• Define calibration gas values to match known values of calibration gas
cylinders (Cal Gas Value).
• Set the pressure and flow of RCU gases without performing a calibra
-
tion or verification (Inject Cal Gas).
• Define cycle times for automatic calibrations or verifications (Set Cal
Timers/Set Auto Timers).
• Set the current date and time (Set Cal Timers/Set Time and Date).
• Temporarily enable or disable auto calibrations or verifications (Set Cal
Timers/Auto Cal On/Off).
CAlIbRATE KEY
Menu Options
Functions/Options
Initiate Cal
Start an automatic, remote, or manual calibration .
Initiate Verify
Start an automatic, remote, or manual verification.
Cal/Verify Data
View latest calibration and verify data.
- Cal data
- Verify data
Cal Gas Values
Enter calibration gas values.
Inject Cal Gas
Set pressure and flow to RCU gases without perform-
ing a calibration.
Set Cal Timers
(applies only for auto cal operations
with RCU, except for setting time
and date)
Set control unit time/date.
Enable or disable auto cal cycles.
Set cal gas times for auto cal/verify operations.
Define recovery durations after a cal/verify.
Configure Verify
Enter verify gas value.
Enter acceptable verify range.
Figure 5-2.
Overview of Calibrate
Key functions.
Содержание Thermox CEM O2/TM
Страница 12: ...xii Thermox CEM O2 TM This page intentionally left blank...
Страница 45: ...Installation 3 23 Figure 3 20 Current output connections example...
Страница 47: ...Installation 3 25 Figure 3 21 Series 2000 standard alarm connections...
Страница 49: ...Installation 3 27 Figure 3 22 Series 2000 controller RCU connections...
Страница 51: ...Installation 3 29 Figure 3 23 Series 2000 2 wire RS 485 communications connections...
Страница 53: ...Installation 3 31 Figure 3 24 Series 2000 4 wire RS 485 communications connections...
Страница 56: ...3 34 Thermox CEM O2 TM This page intentionally left blank...
Страница 116: ...5 26 Thermox CEM O2 TM This page intentionally left blank...
Страница 137: ...Service and Parts 7 3 This page intentionally left blank...
Страница 139: ...Service and Parts 7 5 Figure 7 1 Cell replacement...
Страница 143: ...Service and Parts 7 9 Figure 7 3 Furnace replacement...
Страница 162: ...B 4 Thermox CEM O2 TM This page intentionally left blank...
Страница 166: ...C 4 Thermox CEM O2 TM This page inentionally left blank...
Страница 168: ...D 2 Thermox CEM O2 TM This page intentionally left blank...