4-34 | Thermox CEM O2/TM
Exception log
The exception log provides a history of the following types of events:
System or error messages
Alarm conditions
Hydrocarbon upset condition.
Reset conditions
Start of calibrations or verifications
Accessing the Exception log
Choose Exception Log from the Alarm menu.
The system shows you the most recent exception log entry on line four of
the display, and shows when this event was recorded on line three. Line
three of the display also shows the order in which events are recorded
into the log, with number 1 indicating the most recent entry into the log.
The system is able to store the 20 most recent entries into this exception
log.
Pressing the Down Arrow Key shows later exception log entries. Press-
ing the Up Arrow Key shows more recent exception log entries. Press the
Cancel Key at any time to exit this menu option.
Содержание Thermox CEM O2/TM
Страница 12: ...xii Thermox CEM O2 TM This page intentionally left blank...
Страница 45: ...Installation 3 23 Figure 3 20 Current output connections example...
Страница 47: ...Installation 3 25 Figure 3 21 Series 2000 standard alarm connections...
Страница 49: ...Installation 3 27 Figure 3 22 Series 2000 controller RCU connections...
Страница 51: ...Installation 3 29 Figure 3 23 Series 2000 2 wire RS 485 communications connections...
Страница 53: ...Installation 3 31 Figure 3 24 Series 2000 4 wire RS 485 communications connections...
Страница 56: ...3 34 Thermox CEM O2 TM This page intentionally left blank...
Страница 116: ...5 26 Thermox CEM O2 TM This page intentionally left blank...
Страница 137: ...Service and Parts 7 3 This page intentionally left blank...
Страница 139: ...Service and Parts 7 5 Figure 7 1 Cell replacement...
Страница 143: ...Service and Parts 7 9 Figure 7 3 Furnace replacement...
Страница 162: ...B 4 Thermox CEM O2 TM This page intentionally left blank...
Страница 166: ...C 4 Thermox CEM O2 TM This page inentionally left blank...
Страница 168: ...D 2 Thermox CEM O2 TM This page intentionally left blank...