Serial Communications | A-3
General Serial Communication Commands
This section provides general serial communication commands. Program-
ming commands and Factory setup commands are provided later in this
chapter. Normally, you will use these general serial communications com-
mands.
Read Number (F)
This command allows you to read the value of a variable stored in the
Series 2000 Controller. The data section of this command is a number from
00-FF. These numbers correspond to the type of variable you want to read.
The Variable Table shown in Table B-1 provides a list of these variables and
their location codes. If a variable is selected beyond the end of the Variable
Table shown, an error will be returned. The value returned is followed by
a text string for the units, where applicable.
Example:
Request Oxygen with the following string: >>FEF0536<CR>
Returned string : A20.9 %O2D4
Write Number (h)
This command allows you to set a value at the Series 2000 controller. The
data section of this command is a two digit hex number (00-FF) to point
to the location to write from the variable table (see Table B-1), followed by
the value of the variable. A simple acknowledge is returned if the com-
mand was accepted.
Calibrate (G)
Use this command to start an automatic calibration or verify cycle. The
data section is either:
0 - Calibrate
1 - Verify
2 - Primary cal
Содержание Thermox CEM O2/TM
Страница 12: ...xii Thermox CEM O2 TM This page intentionally left blank...
Страница 45: ...Installation 3 23 Figure 3 20 Current output connections example...
Страница 47: ...Installation 3 25 Figure 3 21 Series 2000 standard alarm connections...
Страница 49: ...Installation 3 27 Figure 3 22 Series 2000 controller RCU connections...
Страница 51: ...Installation 3 29 Figure 3 23 Series 2000 2 wire RS 485 communications connections...
Страница 53: ...Installation 3 31 Figure 3 24 Series 2000 4 wire RS 485 communications connections...
Страница 56: ...3 34 Thermox CEM O2 TM This page intentionally left blank...
Страница 116: ...5 26 Thermox CEM O2 TM This page intentionally left blank...
Страница 137: ...Service and Parts 7 3 This page intentionally left blank...
Страница 139: ...Service and Parts 7 5 Figure 7 1 Cell replacement...
Страница 143: ...Service and Parts 7 9 Figure 7 3 Furnace replacement...
Страница 162: ...B 4 Thermox CEM O2 TM This page intentionally left blank...
Страница 166: ...C 4 Thermox CEM O2 TM This page inentionally left blank...
Страница 168: ...D 2 Thermox CEM O2 TM This page intentionally left blank...