Installation
|
3-11
Weatherproof Wall/Z-Purge Option
Figure 3-10 shows the wall and pole mounting dimensions for the Series
2000 weatherproof Z-purge enclosure. To mount this control unit version,
you must also connect instrument air to the Purge Inlet and set the pres-
sure and flow as indicated on the warning tag on the control unit.
Z-Purge Shutdown Procedure
: Hot internal parts are above the igni-
tion temperature of combustible gases. Remove power from both the
sensor and control unit for 1 hour while maintaining purge air flow
before door is opened, unless the area is demonstrated to be non-haz-
ardous.
Z-Purge Startup Procedure:
Power shall not be restored after the
enclosure has been opened until the enclosure has been purged for 1
hour at a pressure of 0.4” of water.
Don’t open the Z-purge enclosure until you verify that the area has
been classified as non-hazardous.
Pressure
Gauge
1/2” Conduit
Entry (2.22 cm)
Two 3/4” Conduit
Entries (2.70 cm)
Purge Air Inlet
1/8” Compression Fitting
Metering Valve
Exhaust
Restriction
7.50
19.05
5.31
13.49
12.00
30.48
Flange
8.00
20.32
5.88
14.94
Figure 3-10.
Series 2000, weatherproof wall/Z-purge.
Содержание Thermox CEM O2/TM
Страница 12: ...xii Thermox CEM O2 TM This page intentionally left blank...
Страница 45: ...Installation 3 23 Figure 3 20 Current output connections example...
Страница 47: ...Installation 3 25 Figure 3 21 Series 2000 standard alarm connections...
Страница 49: ...Installation 3 27 Figure 3 22 Series 2000 controller RCU connections...
Страница 51: ...Installation 3 29 Figure 3 23 Series 2000 2 wire RS 485 communications connections...
Страница 53: ...Installation 3 31 Figure 3 24 Series 2000 4 wire RS 485 communications connections...
Страница 56: ...3 34 Thermox CEM O2 TM This page intentionally left blank...
Страница 116: ...5 26 Thermox CEM O2 TM This page intentionally left blank...
Страница 137: ...Service and Parts 7 3 This page intentionally left blank...
Страница 139: ...Service and Parts 7 5 Figure 7 1 Cell replacement...
Страница 143: ...Service and Parts 7 9 Figure 7 3 Furnace replacement...
Страница 162: ...B 4 Thermox CEM O2 TM This page intentionally left blank...
Страница 166: ...C 4 Thermox CEM O2 TM This page inentionally left blank...
Страница 168: ...D 2 Thermox CEM O2 TM This page intentionally left blank...