Maintenance and Troubleshooting | 6-7
Thermocouple Failure
This error message occurs when the indicated sensor temperature fall
below -70° C. This error message can be the result of one of the following:
Open thermocouple
Faulty interconnecting wiring
Control unit display module failure
Verify Failure
This error message occurs if the verify gas does not read within the range
you defined. If you receive this message, calibrate the analyzer.
Zero Gas Range Error
This error message occurs if the calibration zero gas does not read within
the software’s allowable limits. Once this error occurs, calibration of the
system is aborted. This error indicates one of the following problems:
Calibration gas setup problem
Incorrect flow rates between calibrated gases
Process pressure incorrectly entered
Cell wiring reserved (Cell + and Cell - wires)
Failed cell (primary calibration should be run at this time)
Leak or plug in plumbing
Содержание Thermox CEM O2/TM
Страница 12: ...xii Thermox CEM O2 TM This page intentionally left blank...
Страница 45: ...Installation 3 23 Figure 3 20 Current output connections example...
Страница 47: ...Installation 3 25 Figure 3 21 Series 2000 standard alarm connections...
Страница 49: ...Installation 3 27 Figure 3 22 Series 2000 controller RCU connections...
Страница 51: ...Installation 3 29 Figure 3 23 Series 2000 2 wire RS 485 communications connections...
Страница 53: ...Installation 3 31 Figure 3 24 Series 2000 4 wire RS 485 communications connections...
Страница 56: ...3 34 Thermox CEM O2 TM This page intentionally left blank...
Страница 116: ...5 26 Thermox CEM O2 TM This page intentionally left blank...
Страница 137: ...Service and Parts 7 3 This page intentionally left blank...
Страница 139: ...Service and Parts 7 5 Figure 7 1 Cell replacement...
Страница 143: ...Service and Parts 7 9 Figure 7 3 Furnace replacement...
Страница 162: ...B 4 Thermox CEM O2 TM This page intentionally left blank...
Страница 166: ...C 4 Thermox CEM O2 TM This page inentionally left blank...
Страница 168: ...D 2 Thermox CEM O2 TM This page intentionally left blank...