Overview | 1-1
OVERVIEW
Sensor Operations
The sample flows through the sensor plumbing because of a convective
motive force. The sample in the vertical call housing is heated to 615°C,
while the sample in the return vertical tube is at about 250°C. This tem-
perature difference generates a density difference, which circulates the
gas through the system.
basic Elements of the Sensor
The CEMO2/TM analyzer consists of the following basic systems:
The Plumbing: All inlet and outlet tubing (cell housing), the sensing
cell and the sensing cell fitting.
The Measuring System: Sensing cell, interconnecting wiring and the
control unit.
The Temperature System: Electrical cell heater (furnace), the type “K”
thermocouple (monitors furnace operating temperature), and the
sensor board containing cold junction thermocouple compensation.
The sensing cell operates at a constant temperature. The circuit board
in the sensor terminal box switches power to the furnace from the AC
mains connected to the sensor.
Содержание Thermox CEM O2/TM
Страница 12: ...xii Thermox CEM O2 TM This page intentionally left blank...
Страница 45: ...Installation 3 23 Figure 3 20 Current output connections example...
Страница 47: ...Installation 3 25 Figure 3 21 Series 2000 standard alarm connections...
Страница 49: ...Installation 3 27 Figure 3 22 Series 2000 controller RCU connections...
Страница 51: ...Installation 3 29 Figure 3 23 Series 2000 2 wire RS 485 communications connections...
Страница 53: ...Installation 3 31 Figure 3 24 Series 2000 4 wire RS 485 communications connections...
Страница 56: ...3 34 Thermox CEM O2 TM This page intentionally left blank...
Страница 116: ...5 26 Thermox CEM O2 TM This page intentionally left blank...
Страница 137: ...Service and Parts 7 3 This page intentionally left blank...
Страница 139: ...Service and Parts 7 5 Figure 7 1 Cell replacement...
Страница 143: ...Service and Parts 7 9 Figure 7 3 Furnace replacement...
Страница 162: ...B 4 Thermox CEM O2 TM This page intentionally left blank...
Страница 166: ...C 4 Thermox CEM O2 TM This page inentionally left blank...
Страница 168: ...D 2 Thermox CEM O2 TM This page intentionally left blank...