3-18 | Thermox CEM O2/TM
Transient and RFI Interference
This section describes transient and RFI interference precautions:
• Although there are transient and noise protectors on all control unit
I/O connections (communications, current outputs, sensor, etc ...), this
protection is intended to act as a last line of defense against unwanted
transient and RFI interference. Proper installation practices to pre-
vent the introduction of transients and noise into the system must
be followed. Inductive loads connected to the control unit must have
transient suppressors installed at the inductive loads. Be sure to place
the transient suppressor as close to the load as possible. Examples of
transient suppressors include MOVs, TRANSORBs, and RC snubbers.
• AC mains supply wiring should not be run in the same conduit with
mains supply wiring that feeds heavy inductive loads.
• Avoid running signal wiring in the same cable or conduit with wires
that power inductive loads unless all the cables within the conduit are
shielded, the inductive loads are small, and transient suppressors are
used at the loads.
• Do not run signal lines in the same cable or conduit with high voltage
lines.
• For optimum noise protection, control unit mains supply wiring
should be connected to a circuit separate from any circuit that could
introduce transients into the system. As an example, do not run mo-
tors, blowers, or air conditioners using the same mains supply circuit
or conduit as the control unit’s mains supply circuit or conduit.
Содержание Thermox CEM O2/TM
Страница 12: ...xii Thermox CEM O2 TM This page intentionally left blank...
Страница 45: ...Installation 3 23 Figure 3 20 Current output connections example...
Страница 47: ...Installation 3 25 Figure 3 21 Series 2000 standard alarm connections...
Страница 49: ...Installation 3 27 Figure 3 22 Series 2000 controller RCU connections...
Страница 51: ...Installation 3 29 Figure 3 23 Series 2000 2 wire RS 485 communications connections...
Страница 53: ...Installation 3 31 Figure 3 24 Series 2000 4 wire RS 485 communications connections...
Страница 56: ...3 34 Thermox CEM O2 TM This page intentionally left blank...
Страница 116: ...5 26 Thermox CEM O2 TM This page intentionally left blank...
Страница 137: ...Service and Parts 7 3 This page intentionally left blank...
Страница 139: ...Service and Parts 7 5 Figure 7 1 Cell replacement...
Страница 143: ...Service and Parts 7 9 Figure 7 3 Furnace replacement...
Страница 162: ...B 4 Thermox CEM O2 TM This page intentionally left blank...
Страница 166: ...C 4 Thermox CEM O2 TM This page inentionally left blank...
Страница 168: ...D 2 Thermox CEM O2 TM This page intentionally left blank...