Maintenance and Troubleshooting | 6-17
If your analyzer doesn’t respond properly to the calibration gas, then it
may indicate a problem with the cell. Before replacing the cell, put the
analyzer through a primary calibration. If this does not work, check the
following:
Check the proper flow rate and pressure of the gases.
Check for leaks or plugged plumbing in the sensor.
Check that the cell is hot enough.
If this does not solve the problem, replace the cell and run another prima-
ry calibration and a regular calibration after the primary calibration.
General Troubleshooting
Your system may pass calibrations yet still seem to be reading incorrect
oxygen levels. If this is the case, check the following.
leak Check
Leaks can lead to inaccurate readings especially if operating under a sig-
nificant pressure or vacuum. Check that all compression fitting and pipe
thread connections are leak tight.
You can check for leaks by sniffing the fittings with another gas. Use a
piece of tygon or plastic tubing with a 1/8” stainless steel nozzle to apply
the gas from a cylinder. Using stainless steel prevents any problems that
might occur with plastic melting the nozzle on hot sensor components.
To do this, connect a vacuum pump to the outlet of the sensor (should be
between 3 and 5 psi) and apply a low calibration gas (we recommend a
low PPM gas for the best results) to the inlet of the sensor. This pulls the
low calibration gas into the sensor. Allow the gas to stabilize and then
measure the cell millivolts. Spray 100% oxygen over the sample inlet
plumbing and when the cell millivolt decrease, it indicates a leak in that
area of the plumbing (a vacuum of the process pulls in the gas). If not con-
venient to view the control unit display, you can also apply a volt meter to
terminals 1 and 2 on the sensor board (Cell + and Cell -) to see if the cell
millivolrs change, indicating a leak.
Содержание Thermox CEM O2/TM
Страница 12: ...xii Thermox CEM O2 TM This page intentionally left blank...
Страница 45: ...Installation 3 23 Figure 3 20 Current output connections example...
Страница 47: ...Installation 3 25 Figure 3 21 Series 2000 standard alarm connections...
Страница 49: ...Installation 3 27 Figure 3 22 Series 2000 controller RCU connections...
Страница 51: ...Installation 3 29 Figure 3 23 Series 2000 2 wire RS 485 communications connections...
Страница 53: ...Installation 3 31 Figure 3 24 Series 2000 4 wire RS 485 communications connections...
Страница 56: ...3 34 Thermox CEM O2 TM This page intentionally left blank...
Страница 116: ...5 26 Thermox CEM O2 TM This page intentionally left blank...
Страница 137: ...Service and Parts 7 3 This page intentionally left blank...
Страница 139: ...Service and Parts 7 5 Figure 7 1 Cell replacement...
Страница 143: ...Service and Parts 7 9 Figure 7 3 Furnace replacement...
Страница 162: ...B 4 Thermox CEM O2 TM This page intentionally left blank...
Страница 166: ...C 4 Thermox CEM O2 TM This page inentionally left blank...
Страница 168: ...D 2 Thermox CEM O2 TM This page intentionally left blank...