Controller / User Interface | 4-31
Relay Configure
All Series 2000 control unit relays are Form A—normally open contacts.
You can, however, use the software to set whether your alarms should
close on an alarm condition (energize on alarm) or open on an alarm
condition (de-energize on alarm). This menu option applies to all alarms,
except for the service and watchdog alarms, which are always set to de-
energize on alarm (fail-safe).
Configuring the relay
1. Select
Relay Configure
from the Alarm menu.
2. Select
De-energize on Alarm
or
Energize on Alarm
.
If you select Energize on Alarm, the following message will appear:
Relays are set to
ENERGIZE on Alarm
If you select De-Energize on Alarm, the following message will appear:
Relays are set to
DE-ENERGIZE on Alarm
Содержание Thermox CEM O2/TM
Страница 12: ...xii Thermox CEM O2 TM This page intentionally left blank...
Страница 45: ...Installation 3 23 Figure 3 20 Current output connections example...
Страница 47: ...Installation 3 25 Figure 3 21 Series 2000 standard alarm connections...
Страница 49: ...Installation 3 27 Figure 3 22 Series 2000 controller RCU connections...
Страница 51: ...Installation 3 29 Figure 3 23 Series 2000 2 wire RS 485 communications connections...
Страница 53: ...Installation 3 31 Figure 3 24 Series 2000 4 wire RS 485 communications connections...
Страница 56: ...3 34 Thermox CEM O2 TM This page intentionally left blank...
Страница 116: ...5 26 Thermox CEM O2 TM This page intentionally left blank...
Страница 137: ...Service and Parts 7 3 This page intentionally left blank...
Страница 139: ...Service and Parts 7 5 Figure 7 1 Cell replacement...
Страница 143: ...Service and Parts 7 9 Figure 7 3 Furnace replacement...
Страница 162: ...B 4 Thermox CEM O2 TM This page intentionally left blank...
Страница 166: ...C 4 Thermox CEM O2 TM This page inentionally left blank...
Страница 168: ...D 2 Thermox CEM O2 TM This page intentionally left blank...