5-12 | Thermox CEM O2/TM
If your O
2
reading on the far left of the display has stabilized, yet is not
near the set point value on the far right of the display, check the verify
gas value you entered and ensure that this value matches the verify gas
cylinder for that gas in the RCU.
When performing a primary calibration or the first calibration after a pri-
mary calibration, the reading and the set point values often won’t match.
Once the first regular calibration after a primary calibration is performed,
the readings should match.
Then the system will display the recovery time remaining:
Recovery xx:xx
System Verifying Cal
After the recovery period, the system returns to process readings.
Manual Verify
1. Enter the verify gas value to match the verify gas you intend to use to
calibrate your system, and also determine the acceptable verify range
(see the “Configure Verify” section later in this chapter for help on
how to do this).
2. Select Initiate Verify from the Calibrate key menu.
3. Select Manual Verify.
You are then prompted to apply the verify gas to the calibration gas inlet
port on the sensor (tubing must be free of oil and dirt):
Inject Span\Vrfy Gas
ENTER to Continue
CANCEL to Abort
4. Press the Enter key once you have applied this verify gas.
Содержание Thermox CEM O2/TM
Страница 12: ...xii Thermox CEM O2 TM This page intentionally left blank...
Страница 45: ...Installation 3 23 Figure 3 20 Current output connections example...
Страница 47: ...Installation 3 25 Figure 3 21 Series 2000 standard alarm connections...
Страница 49: ...Installation 3 27 Figure 3 22 Series 2000 controller RCU connections...
Страница 51: ...Installation 3 29 Figure 3 23 Series 2000 2 wire RS 485 communications connections...
Страница 53: ...Installation 3 31 Figure 3 24 Series 2000 4 wire RS 485 communications connections...
Страница 56: ...3 34 Thermox CEM O2 TM This page intentionally left blank...
Страница 116: ...5 26 Thermox CEM O2 TM This page intentionally left blank...
Страница 137: ...Service and Parts 7 3 This page intentionally left blank...
Страница 139: ...Service and Parts 7 5 Figure 7 1 Cell replacement...
Страница 143: ...Service and Parts 7 9 Figure 7 3 Furnace replacement...
Страница 162: ...B 4 Thermox CEM O2 TM This page intentionally left blank...
Страница 166: ...C 4 Thermox CEM O2 TM This page inentionally left blank...
Страница 168: ...D 2 Thermox CEM O2 TM This page intentionally left blank...